Manufacturers that can quickly explain how they validated their software can clear that part of an FDA inspection or notified body audit in minutes, Johnson & Johnson Vision Care’s Raúl Soto said Oct. 7 at KENX GMP University. Soto shared practical tips on keeping system inventories current, making periodic reviews risk-based, and producing records fast.
A technician monitoring a pharmaceutical production line
Share on LinkedIn

Manufacturers that can clearly explain how they validated their software can often clear that portion of a US Food and Drug Administration (FDA) inspection or notified body audit in minutes, a longtime quality engineer says. Those that can’t should expect investigators to keep digging.

Speaking remotely Oct. 7 at KENX GMP University in Philadelphia, Raúl Soto, Senior Principal Software Quality Engineer at Johnson & Johnson Vision Care, said investigators are quick to read the signals a company sends about its quality culture.

When a quality lead can walk through the validation process with confidence, “your part of the inspection is going to be resolved in five, 10 minutes,” he said. But when records are hard to find or difficult to explain, “a five-minute conversation can turn into a two-hour conversation, easily.”

Soto, who has more than three decades of experience across medical devices, pharma, biotech, and consumer electronics, stressed that his views were his own. His advice, he said, came from practices that “have saved our skin in several inspections, internal and external.”

Start With Inventory

Inspection readiness begins with a controlled, up-to-date inventory of every validated system on site, from production-line controllers and machine vision systems to enterprise software and computerized lab instruments.

Soto recommended keeping it lean: the system name, where it’s used, a contact person, and the latest periodic review. Additional detail only opens the door to more questions, he said.

Keeping the list current matters most. If an investigator picks a system from the list that was retired six months earlier, “you’re not going to get an observation for that,” Soto said. “But they are going to start thinking, ‘Oh, these people are sloppy. I’m going to start looking at this more carefully.’”

“You’re not alone. You have a whole organization that wants you to be successful.” – Raúl Soto

He also cautioned against over-listing. A manufacturing line validated as a single system should appear once, even if it contains 20 separate controllers. Soto recalled a notified body auditor who expected individual inventory entries and periodic reviews for each controller on a line. The team walked the auditor through the logic several times until he understood and accepted it and moved on.

For corporate-owned software a site uses but doesn’t control, such as a shared ERP (Enterprise Resource Planning) system or enterprise lab platform, Soto suggested a separate section listing the responsible corporate contact and at least one backup. Those contacts should be verified quarterly and alerted the moment an inspection begins. Companies shouldn’t wait until an investigator asks about a specific system and “the clock starts running,” he said.

Make Reviews Risk-Based – and Follow Through

Software periodic reviews are a frequent inspection request, Soto said, and many companies make them harder than they need to be. A flat rule that every system gets reviewed every three years may work for a small site, but at a large one it can tie up staff who do “nothing else but reviews.”

He recommended a risk-based schedule, reviewing high-risk systems more often than low-risk ones. The reviews themselves should look at all nonconformances and corrective and preventive actions (CAPAs) tied to a system over the review period, as well as the combined effect of many small changes that individual change control may have missed.

When a review finds a gap, such as a system validated under older procedures that no longer meet current requirements, it should be documented and tracked to closure. The worst outcome is the next review finding that “it’s been two years and we didn’t do anything,” he said.

Soto also warned against committing to a fixed revalidation interval in procedures. “If you put that in your procedure, you’re going to be held to that,” he said, and most companies lack the resources to keep that promise. Instead, he advised tying revalidation to major upgrades, when the system is being rebaselined anyway.

Procedures Should Be Easy To Explain

A Subject Matter Expert (SME) who sits across from an investigator must know the site’s validation procedures “in and out,” Soto said, including how the company decides whether a system needs validation, what documents it requires, and what testing is appropriate. (Related Story: “Confident, Not Just Correct: Readying SMEs for FDA Inspections,” QualityHub, Sept. 25, 2026.)

“That’s usually the main failure nowadays, when people think of documentation as ceremony.” – Raúl Soto

Investigators frequently ask manufacturers to show how their process follows a risk-based approach, and to justify why one system got more documentation than another.

Soto recommended building that logic into procedures visually, using flowcharts and tables tied to recognized frameworks such as GAMP 5 software categories for manufacturing and IT systems, as well as the US Pharmacopeia’s A, B, and C instrument groups for lab equipment.

GAMP 5 is the fifth version of the Good Automated Manufacturing Practice guide published by the International Society for Pharmaceutical Engineering (ISPE), while USP’s General Chapter <1058> for Analytical Instrument Qualification classifies laboratory instruments into three risk-based groups according to their complexity and impact on data quality.

Walking an investigator through a diagram is far easier than through 10 pages of text, and helps turn “two-hour conversations into five-minute conversations,” Soto said.

Find it Fast

Investigators now expect requested documents within roughly 20 minutes to an hour, Soto said, given that most companies use electronic document management systems.

But legacy records can be a stumbling block. He advised companies to scan offsite paper records ahead of time so an unofficial copy can be shown while the original is retrieved, and to ensure at least two trained people still have working access to any retired document or test management system kept in read-only archives.

Documentation isn’t Ceremony

Soto’s sharpest words were aimed at teams that rush validation paperwork to hit deadlines.

“That’s usually the main failure nowadays, when people think of documentation as ceremony,” he said. The cost comes later, during an inspection, when a document “looks like a plate of spaghetti.”

Soto identified three records investigators request most:

  • Validation report: It should tell the full story, including scope changes, deviations from the plan, how defects were resolved, and the release strategy. Any deferred defect needs a documented workaround, or the affected function should be disabled.
  • Traceability matrix: It should link each requirement to the specific test that verified it, and let staff find that evidence in minutes.
  • Periodic review reports: They should show what was found and what was done about it.

The easiest conversation to have with an investigator is “one where you do not say anything and you just point at the page,” Soto said.

Managing the Inspection

Behind the scenes, the team working the inspection “back room” should log every document provided to the investigator, note who discussed it, and keep a copy, Soto said. That team should also have access to every system holding validation records before the investigator arrives, not after.

The back room functions as an operational hub during inspections. It’s where records are gathered, reviewed, analyzed, and prepared before being provided to the FDA, while teams coordinate responses, identify potential issues, prepare SMEs, and manage the flow of information throughout the inspection. (Related Story: “Tips From the ‘Back Room’: 5 Strategies for Successful FDA Inspections,” QualityHub, July 8, 2026.)

And the inspection “front room” is a designated area in a company’s facility where investigators carry out the majority of their work while on site. It’s also where the firm’s representatives respond to the FDA’s questions and provide requested information, among other tasks. (Related Story: “Choosing the Right ‘Front Room’ Host for an FDA Inspection,” QualityHub, Aug. 25, 2026.)

In the front room, Soto urged respect. Investigators “are part of the system that keeps our medical devices and medicines safer,” he said, and if they don’t understand something, it’s the company’s job to explain it better.

Company employees should also answer only what’s asked and resist the urge to fill silences. “If you’re asked what time is it, do not give [the investigator] the story of how to design a clock,” Soto quipped. And if a discussion grows tense, he suggested calling a short break and consulting the back room. (Related Story: “Don’t Say That! How Loose Lips Can Sink Your Next FDA Inspection,” QualityHub, July 8, 2026.)

Finally, Soto said, no one at the table should commit the company to a corrective action without first checking with leadership. “You’re not alone,” he told KENX attendees. “You have a whole organization that wants you to be successful.” (Related Story: “The Most Dangerous Part of an FDA Inspection? The Closeout Meeting,” QualityHub, Sept. 14, 2026.)