Corrective and preventive actions (CAPAs) are only as defensible as the evidence behind them, from the decision to open a CAPA through its final Verification of Effectiveness (VoE) and closure. That evidence should be planned and built into the CAPA process from the start, says a QualityHub expert.
Rebecca Fuller, QualityHub’s SVP of Regulatory Compliance, spoke Oct. 8 at KENX GMP University in Philadelphia. She told pharmaceutical and medical device quality and regulatory professionals that VoE strategy should be considered as early as the problem statement that supports opening the CAPA. The VoE will eventually provide the objective evidence that the action produced the intended, sustainable results.
VoE is the final step of a CAPA, the point at which an organization must demonstrate that its actions corrected the problem and kept it from recurring. Fuller said that step is only as strong as the work that comes before it. A vague problem statement, an incomplete investigation, or an action plan that can’t be traced to a specific cause tends to surface when a team tries to prove its fix worked.
“Oftentimes, companies aren’t thinking about Verification of Effectiveness until that CAPA endpoint, but you must be thinking about it very early on,” Fuller said. “Every failure along the way in your CAPA process is going to impact that Verification of Effectiveness.” (Related Story: “When FDA Smells Blood: Why Risk-Based CAPA Matters Under QMSR,” QualityHub, July 14, 2026.)
“You have to peel that layer of the onion back more and more to really understand what caused [a] direct failure. It may be that there was some human error, absolutely – but you shouldn’t stop there.” – Rebecca Fuller
Device and drug quality frameworks both address VoE directly. ISO 13485:2016, which the US Food and Drug Administration (FDA) incorporates by reference in its Quality Management System Regulation (QMSR), requires medical device manufacturers to verify that a corrective action doesn’t adversely affect the safety and performance of the device or the ability to meet regulatory requirements, and to review the effectiveness of the action taken. For pharmaceutical manufacturers, Sec. 3.2.2 of the ICH Q10 Pharmaceutical Quality System guideline similarly states that “the effectiveness of the [CAPA] actions should be evaluated.”
Fuller described VoE as work that unfolds in stages. It begins before planned actions are implemented, with verification or validation that the action will fix the problem without introducing new risks. Otherwise, a fix can create a new problem that the VoE was never designed to catch. It continues with documented evidence that the actions were actually implemented, and it ends with documented evidence that those actions produced sustainable results in practice, over time.
Because the goal is to show that the problem described in the CAPA no longer exists, VoE activities should be planned around the problem statement. Fuller said successful VoE depends on a good problem statement, a well-understood and actionable root cause, predictable success based on verifying planned actions before implementation, and demonstrable success based on consistent outputs that meet measurable acceptance criteria over time.
Keep Each CAPA Focused
Fuller cautioned that CAPAs that expand to cover every related issue a team uncovers become much harder to verify. When a new problem emerges during an investigation, it often deserves its own record, because proving it was resolved will require a different VoE, she said.
“Don’t be afraid to open another CAPA,” she said, pointing out that a well-structured CAPA links each probable cause to a specific action, ties objective evidence to each action, and measures effectiveness against the original problem statement.
That traceability matters most during an FDA inspection, Fuller said. (Related Story: “5 Pro Tips for a Strong, Vibrant CAPA System,” QualityHub, Dec. 18, 2025.)
Evidence Before Action
QualityHub’s Fuller urged companies to require investigation teams to document what they observe and test to confirm a root cause, rather than simply stating a conclusion.
CAPA forms should require investigation teams to state the “objective evidence of your root cause,” she said. “What did you actually see and do to verify that you truly understand the root cause?”
She also warned against CAPA investigations that stop at the most visible cause.
“You must measure outputs, and figuring out what to measure is sometimes the hardest part.” – Rebecca Fuller
“You have to peel that layer of the onion back more and more to really understand what caused that direct failure,” Fuller said, adding that investigations that default to human error and close with retraining are particularly difficult to verify. “It may be that there was some human error, absolutely – but you shouldn’t stop there.”
Investigation teams also need to record the factors they evaluated and ruled out. “Even if you found that something in particular wasn’t a contributing factor to the problem, you still have to document that you thought through that,” she said.
Implementation Matters but Doesn’t Prove Effectiveness
The most common flaw Fuller sees in VoE plans is that they confirm an action was completed rather than proving it worked. Checking records months later to ensure that a procedure was followed or a setting was correct doesn’t prove effectiveness, she said.
“Is your VoE just verifying that the corrective action was implemented through your quality system? No,” Fuller said. She explained that VoE has to seek out objective evidence and show over time that results consistently meet acceptance criteria.
“You must measure outputs, and figuring out what to measure is sometimes the hardest part,” she said.
Fuller called the failure to plan measurement and data analysis “one of the biggest mistakes” she sees, noting that acceptance criteria are frequently missing from VoE plans altogether.
Setting a Realistic Bar
Fuller said companies that define acceptance criteria often default to zero defects. “They expect to see no further problems at all, but is that realistic?” she asked, explaining that unless the risk truly warrants it, zero is rarely the right target.
Instead, she recommended teams establish three reference points for the defect rate:
- Before the problem occurred;
- While the problem was present; and
- What the manufacturer can reasonably tolerate going forward.
“Was it ever zero? Probably not,” Fuller said, adding that the tolerable rate shouldn’t be a judgment call made at the end of a CAPA. Instead, it should come from approved risk management documentation, such as the process risk assessment, hazard analysis, or DFMEA (Design Failure Mode and Effects Analysis), where the severity and probability of a failure mode have already been evaluated.
“It depends on the severity of the defect when it comes to how much can be tolerated,” she said.
According to Fuller, a defensible VoE plan outlines what will be measured, how often, and how much data is needed. It sets a statistically justified sampling plan, defines a specific acceptance criterion, and explains the risk-based rationale behind it, including which in-process and final inspections would catch the defect if it recurred.
Actions are Meaningless Without Evidence
Fuller noted that many FDA Form 483 observations are issued not because a company did the wrong thing, but because it lacked the objective evidence to show it did the right thing. When that evidence is missing or unclear, “you can’t talk your way around it,” she said.
Reviewers should ask what evidence supports every decision, action, and conclusion in a CAPA, Fuller said. For VoE, that means measurable data drawn from quality system records and attached or clearly referenced so it can be retrieved during audits and inspections. The most convincing data is reproducible and backed by statistical analysis that shows a predefined confidence level was achieved.
Fuller closed by urging quality leaders to challenge CAPAs that lack a credible verification plan before approving them.
“Don’t be afraid to push back,” she said, adding that it’s better to take the time to fix weak plans, decisions, and supporting documentation now than to see them cited in a future inspectional observation.