Risk Comes First: What FDA Wants to See in Your Risk File

Executive Summary

A MedTech manufacturer's risk management file is the first thing FDA investigators will ask to see when performing QMSR inspections, and it needs to reflect real-world complaint data. QualityHub's Rebecca Fuller recommends clinician-approved severity ratings used the same way everywhere, testing plans based on severity, and a matrix that shows how risk drives decisions.
Share on LinkedIn

Medical device makers preparing for a US Food and Drug Administration (FDA) inspection under the Quality Management System Regulation (QMSR) should have their risk management file ready, because it’s fast becoming the first document investigators ask to see.

“Investigators will use the information in that file to pull threads and see how that data has been deployed in different areas of your quality system,” Rebecca Fuller, QualityHub’s SVP of Regulatory Compliance, said at the recent KENX GMP University meeting in Philadelphia.

Put simply, the risk management file is a company’s record of what could go wrong with a device, how badly a patient could be hurt, and what the company is doing about it. The FDA wants to see that information put to work across the business, in design, manufacturing, complaint handling, and corrective and preventive action (CAPA), rather than sitting in a binder.

Starting with the risk file is a change from the era of the FDA’s Quality System Inspection Technique (QSIT), when facility inspections typically opened with areas such as CAPA, management controls, and design controls. The agency retired QSIT in February 2026 when the QMSR came into force. (Related Story: “R.I.P. QSIT (1999-2026): A Requiem for FDA’s Inspection Technique,” QualityHub, Jan. 27, 2026.)

The ground rules for this work come from ISO 14971:2019 – the international standard for medical device risk management – and in practice, device makers can’t ignore it, Fuller said. That’s because the QMSR incorporates the ISO 13485:2016 quality system standard by reference, and ISO 13485 points companies to ISO 14971. (Related Story: “Risk Management & QMSR: 7 Questions Answered By FDA,” QualityHub, Jan. 15, 2026.)

A File That Has to Keep Up

The most common problem Fuller sees is a risk file that’s built during product development and then left alone. Once a device is on the market, complaints may reveal new ways it can fail, but those failures don’t always make their way back into the file.

When she audits a company’s quality system, Fuller compares its complaint records with its risk file. For each reported failure, she asks where it appears in the file and how serious it’s rated. Roughly half the time, she said, the failure isn’t in the file at all.

Her advice is to run risk management like any other quality process, with clear procedures, trained owners, and a steady flow of manufacturing and complaint data feeding back into the file.

One Harm, One Severity Rating

At the heart of any risk file is severity, a rating of how badly a patient could be hurt if something goes wrong. Fuller says that rating should be set once, in a master list of possible harms, and then used consistently everywhere else, from early design reviews to the detailed failure analyses engineers perform on the device and its manufacturing process.

That master list should be led and signed off by physicians and clinical staff, not engineers, because judging how badly a failure could hurt a patient is “a clinical decision, not an engineering one,” Fuller said. Once it’s approved, a severity rating should change only if real-world events show that patients were hurt worse than predicted.

“Risk management isn’t a file you finish and put on a shelf. It’s how you make decisions.” – Rebecca Fuller

Problems start when teams work separately. Software developers or manufacturing engineers sometimes assign their own severity ratings without checking the master list, leaving the company with conflicting numbers for the same harm. Fuller noted that a glitch on the production line doesn’t hurt anyone by itself, but it does create a defect that may eventually reach a patient, and the severity of that harm has already been decided.

“We don’t have a harm until we get to the patient,” she said.

The same rule applies when another company builds the device. The device maker owns the master harms list and should make sure its contract manufacturer uses the same severity ratings, ideally with that requirement written into their quality agreement. (Related Story: “FDA Investigator Reminds MedTechs: Monitoring of Suppliers Must Include Risk Considerations,” QualityHub, May 7, 2025.)

Plan With Severity, React With Probability

Many companies score risk by multiplying severity by probability, or how likely a failure is to happen. The result is called a risk priority number, or RPN. Fuller finds that some manufacturers lean too heavily on that number.

When a company is planning ahead, such as deciding how many units to test, how thoroughly to validate a manufacturing process, or how closely to watch a supplier, Fuller recommends using severity alone. The more serious the potential harm, the more testing and oversight it gets.

The reason for leaving probability out is that, before a product reaches the market, probability is only a guess.

When a company sets up its own product testing plans, Fuller said, it doesn’t know the real probability of a failure. All it has is the estimate its engineers predicted during development.

Those predictions tend to be optimistic, so building them in usually means less testing. If real-world data later shows a failure happens more often than expected, the company may have to redo costly validation work. Testing more up front costs a little more but builds in a cushion.

The added cost is confined to a few validation batches, Fuller said, and then “you’re going to have yourself well protected. Even if the probability changes, it’s not going to matter.”

Probability earns its place once something has already gone wrong, such as a rise in complaints or a patient injury. At that point the company has real data, and it should compare how often the problem is actually happening with what the risk file predicted. That comparison helps decide whether an investigation, CAPA, or product recall is needed. (Related Story: “When FDA Smells Blood: Why Risk-Based CAPA Matters Under QMSR,” QualityHub, July 14, 2026.)

Fuller also urged companies to keep their risk ratings simple. She favors a five-point scale over a 10-point one. The larger scale invites long debates over small differences that rarely change the outcome.

“Who’s going to argue six versus five, or seven versus eight?” Fuller asked.

Risk assessment is “an exercise in relativity,” she said. The goal is to rank risks against one another so a company knows where to focus its money and effort, not to produce perfect numbers.

Show How Risk Drives Decisions

To show FDA investigators exactly where risk drives decisions, Fuller recommends a risk decision matrix, often kept as a spreadsheet. It lists each area of the quality system, the decisions there that depend on risk, what triggers them, which risk information is used, and who’s responsible.

For example, in complaint handling, each complaint category can link to a severity rating in the risk file, which then sets how deeply the complaint is investigated. A death or serious injury goes straight to a Health Hazard Evaluation, or HHE, in which the company quickly weighs how dangerous the problem is and whether products need to come off the market.

With that risk decision matrix in hand, an investigator can check a few examples, confirm the company follows its own rules, and move on. Without it, a company is left answering an open-ended question about how it uses risk, and the answer is rarely as clear as it should be.

“Risk management isn’t a file you finish and put on a shelf. It’s how you make decisions,” Fuller said. “When an investigator can follow a single harm from your master list all the way to your complaint files and see the same severity at every step, you’ve shown them a system that works.”